Prius Healtcare Enhance DX Features
- The Enhance DX employs 8″ air bladders with the cell in cell design (4.5″ of therapy, 3.5″ of static air), to provide therapeutic alternating pressure and also allow the mattress to stay inflated in a power outage
- Durable woven polyester bottom cover with PVC lining for infection control
- The Enhance DX replacement mattress can accommodate patient weights up to 500 lbs
- Low friction, low shear cover
- Easy to clean and maintain
- 2 in 1 alternation
- Pressure mapping
Prius Healthcare Master Control Unit Features
- The electronic control provides 8 adjustable levels of comfort depending on the patient’s therapy requirement
- 2-in-1 alternating air cycle ensures the maximum protection
- The power failure alarm is triggered during a power outage
- The Low-Pressure Alarm is triggered when the air pressure of the mattress is lower than the default setup
What to buy with Pirus Enhance DX Mattress?
- Medline ThermaLux Spread Blanket
- Fabrication Blanket Lift Bar
- Essential Medical Tilt Top Overbed Table
How to use Prius Healthcare Enhance DX Mattress System?
- Unpack the system and place the pump at the foot end of the bed.
- Remove the existing mattress from the bed frame.
- Place the Enhance DX mattress on the frame and position mattress so the air tube is at the foot of the bed.
- Secure the Enhance DX mattress straps to the bed frame.
- Hang the control unit on the footboard of the bed frame.
- Connect the mattress air hose to pump
- Check air hoses under the mattress to make sure they are not kinked.
- Plug in the control unit and turn it on.
- The pump will now inflate the mattress.
- Press the control button on the pump to select alternation or static therapy.
- Set the required pressure by pressing the “+” to increase pressure or to reduce pressure.
- Pressure is constantly monitored by the control unit. When pressure is less than the selected pressure the “Low Pressure” LED will illuminate and an alarm will sound.
- For emergency, mattress deflation turns the CPR valve to the ‘’OPEN’’ position. The mattress should deflate within 20 seconds. To close the CPR simply turn the valve to the ‘’CLOSE’’ position.
How to Clean Enhance DX Mattress System?
- The Mattress
The mattress should be cleaned weekly using a damp soft cloth and mild detergent. The mattress cover can be washed and thermally disinfected in a washing machine by following the procedure below. (Do not use a phenol-based cleaning solution)
- The Master Control Unit
The pump unit should also be cleaned weekly using a damp soft cloth and mild detergent. The pump casing is manufactured from ABS plastic. If the case is soiled the pump can be wiped down with a solution of sodium hypochlorite dilution, 1000ppm, or any EPA approved hospital grade disinfectant. (Do not use a phenol-based cleaning solution.)
- The air filter
It should also be cleaned and checked at a minimum of every six months. The air filter can be removed by removing filter cover and pinching center of the filter and pulling outward from the back of the cover.
Storage and Care
- Master Control Unit
- Check the power cord and plug for abrasions and excessive wear
- Plug in the unit and verify air flow from the hose connection ports.
- Place in plastic bag for storage
- Mattress
- Check the air manifold for kinks or breaks and replace if necessary
- Set CPR valve to “OPEN” and disconnect the air hose to the pump. The mattress will now deflate and can be packed for storage.