Product Description
Verify SixCess Steam 20/Bx
Steris LCC003″ title=”Indicator Emulating Challenge Pack
Verify SixCess Steam 20/Bx
Steris LCC003″ />Indicator Emulating Challenge Pack
Verify SixCess Steam 20/Bx
Steris LCC003
Manufacturer LCC003Brand VERIFY™ SixCess®Manufacturer SterisApplication Sterilizer Monitoring Service Challenge PackClass Class 6Indicator Change Yellow to Blue/PurpleSterilization Compatibility SteamTemperature 270°F (132°C)Time 4 Minute Processing TimeUNSPSC Code 42281810Features
Challenge Packs provide consistent tests of the ability to penetrate the load with sterilant Each challenge pack represents a difficult challenge to the sterilization process and provides assurance that the load may be released for use Contain a Class 6 emulating indicator within a specially designed barrier system Offer a fast, accurate and cost effective means of sterility assurance monitoring and release of steam sterilization loads Not a substitute for biological indicators and are not intended to release loads containing implants Monitor the cycle specific sterilization parameters required Indicator ink performance meets ANSI/AAMI/ISO 11140-1 Class 6 Emulating Indicator criteria Immediately identifies failed sterilization cycles Provides a cost-effective means to monitor every sterilization load Products for use in extended cycle times